Cooling bath

A cooling bath or ice bath, in laboratory chemistry practice, is a liquid mixture which is used to maintain low temperatures, typically between 13 °C and −196 °C. These low temperatures are used to collect liquids after distillation, to remove solvents using a rotary evaporator, or to perform a chemical reaction below room temperature (see Kinetic control).

Cooling baths are generally one of two types: (a) a cold fluid (particularly liquid nitrogen, water, or even air) — but most commonly the term refers to (b) a mixture of 3 components: (1) a cooling agent (such as dry ice or ice); (2) a liquid "carrier" (such as liquid water, ethylene glycol, acetone, etc.), which transfers heat between the bath and the vessel; (3) an additive to depress the melting point of the solid/liquid system.

A familiar example of this is the use of an ice/rock-salt mixture to freeze ice cream. Adding salt lowers the freezing temperature of water, lowering the minimum temperature attainable with only ice.

Mixed solvent cooling baths (% by volume)
 % Glycol in EtOH Temp (°C)  % H2O in MeOH Temp (°C)
0% −78 0% −97.6
10% −76 14% −128
20% −72 20% N/A
30% −66 30% −72
40% −60 40% −64
50% −52 50% −47
60% −41 60% −36
70% −32 70% −20
80% −28 80% −12.5
90% −21 90% −5.5
100% −17 100% 0
This article is issued from Wikipedia. The text is licensed under Creative Commons - Attribution - Sharealike. Additional terms may apply for the media files.